Clinical trial · Interventional
CT-validation of Ultrasound Based Planning for High Dose Rate (HDR) Prostate Brachytherapy Using Vitesse
High Dose Rate (HDR)Prostate Brachytherapy With Vitesse tm
- Source
- ClinicalTrials.gov
- Retrieved
- Sep 8, 2026
- Layer
- normalized (units and labels harmonized; values unchanged)
- Run
- ING-CLINICALTRIALS-20260908-000001
Summary
Brief summary (as posted)
High Dose Rate (HDR) temporary prostate brachytherapy offers a precise form of dose escalation for prostate cancer. Needles are placed, the position is confirmed and treatment is delivered. Previously High Dose Rate prostate brachytherapy was performed under TransRectal UltraSound (TRUS) guidance but planned with Computed Tomography (CT) imaging which introduced a source of error through needle displacement while transporting and repositioning patient. Recently Varian has introduced a one-step procedure where both implant and planning are based on UltraSound imaging and performed without patient repositioning. This planning system is approved by Health Canada and is commercially available. This study will use High Dose Rate brachytherapy to treat 20 men. Treatment will be planned with TransRectal UltraSound and validated using Computed Tomography imaging.
Conditions
Conditions (4)
Free-text conditions as registered, with the CancerIndex entity they were reconciled to and the match type.
| Condition (as posted) | Mapped entity | Match | Confidence |
|---|---|---|---|
| High Dose Rate Brachytherapy | — | UNRESOLVED | — |
| Interstitial Radiation | — | UNRESOLVED | — |
| Prostate Cancer | Malignant Prostate Neoplasm | CURATED_EXACT | 0.92 |
| Ultrasound-based Planning Compared to CT-based Planning | — | UNRESOLVED | — |
Interventions
Interventions (1)
| Intervention | Type | Mapped drug | Match |
|---|---|---|---|
| HDR prostate brachytherapy | Radiation | — | UNRESOLVED |
Design
Arms and outcomes
Arms (1)
- type
- EXPERIMENTAL
- label
- HDR prostate brachytherapy
- description
- An array of 17 gauge steel needles, 20 cm in length, is inserted under Ultrasound (US) guidance into the prostate and advanced to the base of the prostate. A template is used to maintain spacing and parallelism and the needles are individually locked into the template once positioned. A continuously acquired set of US images is obtained for treatment planning purposes. Each needle is connected to a Varisource afterloader and treatment is delivered using a 10 Curie Iridium-192 source. The needles are then removed.
- interventionNames
- Radiation: HDR prostate brachytherapy
Primary outcomes (1)
- measure
- Accuracy of Needle Tip Identification on TransRectal UltraSound Compared to Computed Tomography
- timeFrame
- Treatment was planned and delivered based on ultrasound identification of needles and needle tips relative to prostate. All measurements and comparisons of measurements pertain to the day of the procedure.
- description
- Precise location of the tip of each needle in 3-dimensional space relative to the prostate determines the pattern of dose delivery. Inaccuracies in needle tip location may lead to differences detected between UltraSound(US)-based and Computed Tomography (CT)-based plans.The distance of the needle tip (as identified on either CT images or US images) protruding beyond the template was measured in mm. The template acted as a fixed reference point as it was locked into the stabilizer/stepper apparatus. The 2 protrusion readings for each needle tip were compared, the difference recorded in mm, and categorized as \< 1 mm, 1-2 mm, 2-3 mm and \> 3 mm.
Eligibility
Eligibility (as posted)
- Sex
- Male
- Minimum age
- 45 Years
- Maximum age
- 80 Years
Show eligibility criteria text
Inclusion Criteria: 1. Patients must have histologically proven adenocarcinoma of the prostate. 2. Patients must have intermediate risk prostate cancer. (Clinical stage ≤ T2c, Gleason score = 7 and initial prostate specific antigen (iPSA) ≤ 20, or Gleason score ≤ 6 and iPSA \> 10 and ≤ 20. 3. Patients must be fit for general or spinal anaesthetic. 4. Patients must have an estimated life expectancy of at least 10 years. 5. Patients must have an Eastern Cooperative Oncology Group (ECOG) performance status of 0 - 2. 6. Patients must have no contraindications to interstitial prostate brachytherapy. 7. Patients on coumadin therapy must be able to stop therapy safely for at least 7 days. Exclusion Criteria: * Men not able to fully understand the trial and the informed consent document * Men suffering from claustrophobia and unable to have a Computed Tomography scan * Men not wishing to have a cone-beam Computed Tomography scan following the insertion of the High Dose Rate brachytherapy needles to verify the accuracy of the treatment plan * Men who cannot safely discontinue blood thinners for a few days.